Compliance to the Electromagnetic Emissions
TheDSL-8001 system is intended for use in the electromagnetic environment specified below.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF Emissions
CISPR 11 (FA)
EN 55011 (CE)
Group 1
The equipment uses RF energy that is necessary for the
internal functioning of the equipment itself. Therefore, its
RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR 11 (FA)
EN 55011 (CE)
Class A
Harmonic
Emissions
IEC 61000-3-2
NA (FA)
Class A
(CE)
Voltage
Fluctuations/
Flicker Emissions
IEC 61000-3-3
NA (FA)
Complies
(CE)
This equipment is suitable for use in all establishments
other than domestic and those directly connected to a
low-voltage power supply network which supplies buildings
used for domestic purposes.
Compliance to the Electromagnetic Immunity (1)
The DSL-8001 system is intended for use in the electromagnetic environment specified below. It is
recommended that the DSL-8001 system be used in such an environment.
Immunity Test
IEC60601-1-2
Test Level
Compliance Level
Electromagnetic Environment -
Guidance
Electrostatic
Discharge (ESD)
IEC61000-4-2
±
2, 4, 6kV
Contact
±
2, 4, 8kV Air
±
2, 4, 6kV Contact
±
2, 4, 8kV Air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient / burst
IEC61000-4-4
±
2kV for power
supply lines
±
1kV for
input/output lines
±
2kV for power
supply lines
±
1kV for
input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC61000-4-5
±
0.5, 1kV:
differential mode
±
0.5, 1, 2kV:
common mode
±
0.5, 1kV:
differential mode
±
0.5, 1, 2kV:
common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips,
short
interruptions,
and voltage
variations on
power supply
input lines.
IEC61000-4-11
for 0.5 cycles
<5% U
T
(>95% dip in U
T
)
for 5 cycles
40% U
T
(60% dip in U
T
)
for 25 cycles
70% U
T
(30% dip in U
T
)
for 5 cycles
<5% U
T
(95% dip in U
T
)
for 0.5 cycles
<5% U
T
(>95% dip in U
T
)
for 5 cycles
40% U
T
(60% dip in U
T
)
for 25 cycles
70% U
T
(30% dip in U
T
)
for 5 cycles
<5% U
T
(95% dip in U
T
)
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the DSL-8001 requires
continued operation during
power mains interruptions, it is
recommended that the
DSL-8001 is equipped with an
internal battery (option) or is
powered from an uninterruptible
power supply.
Power
Frequency
(50/60Hz)
Magnetic Field
IEC61000-4-8
3A/m
3A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
Note : U
T
is the AC mains voltage prior to application of the test level.
xxii
Summary of Contents for DSL-8001
Page 4: ...Blank Page ...
Page 39: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Page 44: ...2 6 Blank Page ...
Page 60: ...3 16 Blank Page ...
Page 61: ...Wiring Diagram Chapter 4 Block Wiring Diagram 4 2 4 1 ...
Page 64: ...4 4 Blank Page ...
Page 72: ...6 6 Blank Page ...
Page 102: ...7 30 Blank Page ...
Page 148: ...11 6 Blank Page ...