Rapid Exhaust Test
Exhaust control of the rapid exhaust valve is tested.
If the test result is “NG”, malfunction of the NIBP unit rapid exhaust valve can be considered.
1. Select the “NIBP Maintenance” in the “Maintenance” to display the “NIBP Maintenance”
window.
2. Select "Rapid Exhaust Test" under the "NIBP Maintenance" window.
3. Select “START/STOP”, then the LED on the key will light in green and the rapid exhaust test
will be performed.
4. When the rapid exhaust test is completed, the LED on the key will be white and the test
results will be displayed.
Criteria : Rapid exhaust speed should be within 10 sec.
If the rapid exhaust does not meet the criteria, a malfunction of the rapid exhaust
valve on the NIBP unit can be considered. Replace the NIBP unit.
9
-
11