Inflation Test
Pressure performance of the NIBP pump is tested.
If the test result is “NG”, malfunction of the NIBP unit pump can be considered.
1. Select the “NIBP Maintenance” in the “Maintenance” to display the “NIBP Maintenance”
window.
2. Select "Inflation Test" under the "NIBP Maintenance" window.
3. Select “START/STOP”, then the LED on the key will light in green and the Inflation Test will be
performed.
4. When the Inflation Test is completed, the LED on the key will be white and the test results will
be displayed.
Criteria : The inflation between 0 to 300mmHg should be within 6 seconds.
If the inflation test does not meet the criteria, a malfunction of the pump on the
NIBP module can be considered.
Replace the NIBP unit.