Check if the corresponding port (COM1 or COM2) is set for the TCON on the
“Communication Setup” screen.
The <TCON Cross Talk> message is displayed.
Cause
: Other bedside monitor with the same TCON ID may exist.
Solution : Check the TCON ID of other bedside monitor in the same TCON group, and if the
same TCON ID exists, set a different TCON ID.
<TCON Invalid Setting> message is displayed.
Cause
: The TCON channel setting is outside the allowable range.
Solution : Make sure the TCON channel number is within the allowable range on “Initial Setup”
> “TCON Setup” > “Channel”.
Battery
The battery cannot be charged. Battery discharges quickly after charging and does not
operate.
Cause 1 : The battery life has expired.
Solution : The battery is a consumable product. Replace it every year.
Cause 2 : The ambient temperature is too high or too low.
Solution : Charge the battery in an ambient temperature between 10 to 30ºC.
Cause 3 : The AC power is not ON.
Solution : Connect the power cable.
The operation time is short although the battery is charged.
Cause 1 : Use in a location that has a lower ambient temperature.
Solution : If the battery is used in a particularly cold location, the operation time becomes
shorter.
Cause 2 : The battery life has expired.
Solution : The battery is a consumable product. Replace it every year.
The battery LED does not light during battery charge.
Cause
: The battery temperature is high.
Solution : For safety reasons, if the battery is charged in a repetitive manner, its temperature
will be high and the charging operation will be in a standby mode until the battery
temperature decreases. When the battery temperature decreases, the charging
operation will automatically start.
The <Charge Battery> message is displayed.
Cause
: Due to low battery condition, the operable time is less than 5 minutes.
Solution : Plug in the AC power cable.
The <Low Battery> message is displayed.
Cause
: Due to low battery condition, the operable time is less than 15 minutes.
Solution : Plug in the AC power cable.
W A R N I N G
The
users should not perform the battery installation procedure as electric
shock may result to the patient and operator or malfunction of the
equipment may occur.
8
-
11
Summary of Contents for DSL-8001
Page 4: ...Blank Page ...
Page 39: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Page 44: ...2 6 Blank Page ...
Page 60: ...3 16 Blank Page ...
Page 61: ...Wiring Diagram Chapter 4 Block Wiring Diagram 4 2 4 1 ...
Page 64: ...4 4 Blank Page ...
Page 72: ...6 6 Blank Page ...
Page 102: ...7 30 Blank Page ...
Page 148: ...11 6 Blank Page ...