![Fukuda Denshi DS-8200 Service Manual Download Page 197](http://html.mh-extra.com/html/fukuda-denshi/ds-8200/ds-8200_service-manual_2344851197.webp)
Replacing the Consumable Parts
This section describes the disassembly and assembly procedures of the unit for consumable parts
replacement.
●
Consumable Parts
The consumable parts are as follows.
Item Name
Part
Number
Usable Life
Cable with Lithium Battery
(Long-Term Backup Battery)
1M010480A
6 years (It may decrease depending on the
frequency and the environment of usage.)
6 years (It may decrease depending on the
frequency and the environment of usage.)
AC power supply unit
SNP-G125-MF
1T010107A
6 years (It may decrease depending on the
frequency and the environment of usage.)
LCD Touch Panel
9D0113130
●
Check Procedure Before Replacement
The following procedures must be performed before replacement.
・
Verify that the system starts properly.
・
Back up the setup data and patient data in a CF card as a measure of precaution.
●
Procedure Before Disassembly
C A U T I O N
This procedure must be performed to prevent electric shock and to
prevent damage/deterioration of the electronic parts resulting from
electrostatic discharge.
・
Disconnect the power supply cable.
・
Wear electrostatic strap, etc.
Replacing the Long-Term Backup Battery
●
Usable Life of the Long-Term Backup Battery
The lithium battery (BR3032) is used for the long-term backup battery and the usable life is
approximately 6 years although it may change depending on the frequency of usage. When using
the unit beyond the usable life, it may be unable to backup the data such as display setup when
the power is turned off.
Replace the long-term backup battery periodically to maintain normal function.
C A U T I O N
z
The setup data is initialized when replacing the battery. Back up the
setup data in such as CF card before replacing.
z
Use the specified backup battery when replacing. If unspecified
backup battery is replaced, it may cause leakage, heating and
explosion.
z
Do not peel off or damage the exterior tube. It may short out and
cause leakage, heating and explosion.
●
Tools for Replacement
・
Phillips screwdriver (Large)
・
Pliers
N O T E
z
Prepare the “Cable with lithium battery (1M010480A)” when replacing.
10-31
Summary of Contents for DS-8200
Page 4: ...Blank Page ...
Page 44: ...1 4 Blank Page ...
Page 112: ...4 40 Blank Page ...
Page 116: ...5 4 Blank Page ...
Page 156: ...8 4 Blank Page ...
Page 166: ...9 10 Blank Page ...
Page 208: ...11 4 ...