4
Press again the [Key Lock] key for more than 2 seconds.
The message will disappear and the keys will be activated again.
C A U T I O N
z
If stains cannot be removed from the touch panel surface, wipe softly with
a dry or ethanol dampened cleaning cloth. Never use strong-acidic
cleaning solution.
z
A special coating is applied to the surface of the touch panel. Do not wipe
the surface with a cloth or gauze with coarse texture. Wipe the surface with
the soft cleaning cloth provided as optional accessory or with an eyeglass
cleaning cloth.
Cleaning the Housing
Clean the housing using tightly squeezed gauze or an absorbent cotton cloth dampened with
alcohol or a neutral cleanser.
C A U T I O N
z
Clean the equipment frequently so stains can be removed easily.
z
To prevent injury, it is recommended to wear gloves when cleaning the
equipment.
z
Do not allow liquids or cleaning solution to enter the monitor or connectors.
z
Do not use organic solvents, thinner, toluene and benzene to avoid
damaging the resin case.
z
Do not polish the housing with abrasive or chemical cleaner.
z
When sterilizing the entire room using a spray solution, pay close attention
not to have liquids get into the monitor or connectors.
z
Use only neutral detergent to clean the housing. Do not use chemical cloth,
scrub brush, abrasive, polishing powder, hot water, volatile solvent and
chemicals (cleanser, thinner, benzine, benzol, and synthetic detergent for
house and furniture), or sharp-edged tools. The surface coating may be
damaged, resulting in discoloration, scratches, and other problems.
Cleaning the ECG Lead Cable, Relay Cable
y
Do not use thinner, toluene, or other organic solvents to clean the cables.
y
After using the cables, clean with neutral detergent or 70% isopropyl alcohol. Do not pull the
cable and do not hold the connector part when cleaning.
(It may degrade the cable coating and result in damage. Particularly organic solvents and
antiseptic solution such as cresol soap solution will degrade the cable coating.)
y
After cleaning, dry it completely before usage.
y
Do not use high temperature sterilization such as steam or EOG method.
y
Do not soak the cable in water or antiseptic solution.
Disinfecting the Blood Pressure Transducers
Disinfect the blood pressure transducers according to the manufacturer’s guidelines.
Do not soak the cable in water or antiseptic solution.
Cleaning and Disinfecting the Temperature Probe
y
Disinfect the temperature probe according to the guidelines provided with the product.
y
When cleaning, follow the procedure below.
(1) Wipe the probe using 70% isopropyl alcohol in cotton.
(2) Dry it completely to air before use
y
Do not soak the cable in water or antiseptic solution.
10-24