2
Calibration
This section describes CO
2
calibration of the HCP-800/810.
The HCP-800/810 CO
2
calibration is operated on the display of the DS-8200 system main unit.
CO
2
Calibration Step
Perform CO
2
calibration using Calibration Gas (Calibration Gas Kit 03046530RFBD).
N O T E
Calibrate after 5 minutes or more after the power of HCP-800/810 is turned ON.
1
Press the [Menu]
→
[CO
2
]
→
[CO
2
Calibration] keys.
2
Press the [Start Cal] key after the “Calibration Ready” is displayed.
3
When “Feed CAL. GAS” is displayed, keep on injecting the Calibration Gas continually until
the message disappears.
10-20