NIBP Test
For details, refer to “Chapter 9 Maintenance” of “Super Unit HS-8000 Series Service Manual”.
●
Software Dip-SW
Sets up each software Dip-SW.
VF Detection (Sensitive)
[OFF]: There is less error detection of VF on normal waveform than the existing frequency analysis
method.
[ON]: VF is detected using both baseline measurement method and frequency indicator.
VT/RUN Detection (Sensitive)
[OFF]: Transient tachycardia is not detected.
[ON]: Transient tachycardia is detected as the continuous VPC rather than the average heart rate.
SpO
2
Sync. Tone Level Difference
Set the SpO
2
PR Synchronized Tone using the [Low] and [High] keys.
Display Menu on External Monitor
[OFF]: The menu is not displayed on the external monitor.
[ON]: The menu is displayed on the external monitor.
●
Analogue Adjust
For details, refer to “Chapter 9 Maintenance” of “Super Unit HS-8000 Series Service Manual”.
10-19