TCON Sens. Adjustment (Only for USA)
Adjusts the sensitivity of TCON.
The fine adjustment can be performed with “+” and “-“ keys.
●
Printer Adjustment
Printer
Status
Monitors opening and closing of cassette, recording papers and thermal head temperature.
・
Cassette
Status Information
Set Nil
Cassette is closed.
Cassette is open.
・
Printer Paper
Status Information
Set Nil
Recording paper is set.
Recording paper is unset.
・
Thermal Head Temperature
Temperature displays the resolution by 1
℃
unit.
Status Information
Normal Abnormal
Thermal head temperature
is less than 70
℃
.
Thermal head temperature exceeds 70
℃
. Lower the
temperature below 60
℃
to return to normal.
Printing on and feeding the paper cannot be performed.