AD
This screen displays the cuff pressure waveform and pulse waveform during NIBP
measurement.
1 Apply the cuff to the arm in a same condition as performing the normal
measurement.
2 Display the “AD” screen, and press the “NIBP Start/Stop” key.
3 The NIBP measurement will start, and cuff pressure waveform and pulse
waveform will be displayed.
4 Press the
key to return to the “NIBP Meas” screen.
Red: Cuff Pressure Waveform (Pressure Sensor 1)
Green: Cuff Pressure Waveform (Pressure Sensor 2)
Yellow: Pulse Waveform (OSC1)
Blue: Pulse Waveform (OSC2)