Channel ID / Group ID Setup
1 Select
è
è
.
The telemeter setup menu will be displayed.
2 To set the telemetry channel ID, a password must be input. First, enter the
password.
Enter the 4-digit password, “7001” using the
keys.
3 Enter the channel ID.
Enter the 4-digit medical telemetry channel ID. The channel
ID other than medical telemetry will not be set.
Reference
【
Adjustable Range of Telemetry Channel
】
ch3001
~
ch6979
(
SS
80
~
SS
99: not adjustable)
4 Enter the group ID.
Enter the group ID in the range from 00 to 60.
5 Save the channel ID and group ID.
Select
, and press
.
NOTE
When the monitor indicates that the measurement data is out of range
(“
´´´
” display), the minimum or maximum value of the range will be displayed
at the telemetry center.
【
Out of Range
】
【
Telemetry Center
】
HR
301bpm or above 300bpm
RR
151bpm or above 150bpm
3
-