158
Maintenance Menu
4.
Replace Electrodes
It is recommended that the electrodes be replaced after 2,500 arc discharges. When
the number of arc discharges reaches a count of 3,500, a message prompting to
replace the electrodes is displayed immediately after tuning on the power. Using the
electrodes without a replacement will result in greater splice loss and reduced splice
strength. In certain high strength applications, cleaning the electrodes can be
beneficial. Contact your local authorized distributor for information on these
applications.
Replacement Procedure
(1)
Execute [Replace Electrodes] in
[Maintenance Menu].
(2)
Instruction messages will appear on the
screen to turn off the power. Press and
hold ENT key till the LED color
changes from green to red.
(3)
Remove the old electrodes.
1.
Remove the electrode cover.
2.
Loosen the screw located on
electrode fixture. Take out the
electrode.
3.
Clean the new electrodes with alcohol-moistened clean gauze or lint-free
tissue and install them in the splicer.
4.
Fit the electrode in the electrode fixture.
5.
Place the electrode fixture on the splicer and tighten screw.
6.
Place the electrode cover.
(4)
Turn on the power, prepare and load fibers into the splicer and press ENT
key. After executing the arc calibration, the splicer will repeat the arc
discharge in succession to stabilize the electrodes.
Upon completion of the arc discharge process, the splicer executes arc
calibration again. The operator should repeat arc calibration until the
“Test Finish” message appears. For details of the arc calibration, see
section: [Arc Calibration]
• The message of “Replace Electrodes” is displayed when the arc count
exceeds the setting threshold value. This threshold value can be changed.
• Do not pull out wiring when replacing electrode.
• To secure the electrodes, tighten screw while pushing the
electrode collar to the electrode fixed base.