Maintenance Menu
1.
Arc Calibration
(1)
Select [Arc Calibration] in [Maintenance Menu] to display [Arc Calibration]
screen.
(2)
Set prepared fibers in the splicer.
See section [
• Use standard SM or DS fiber for arc calibration.
• Glass clamping is strongly recommended for arc calibration. The error
margin of the calibration value may be larger in case of the coating
clamping.
• The error margin of the calibration value may be larger when the surface
of the fiber is dusty. Set the cleaved fiber again if the surface of the fiber is
dusty.