Turning splicer ON
Confirming splice, heater and operation modes
Cleaning coating or sheath of fiber
Placing protection sleeve over fiber
Cleaning fiber
Cleaving fiber
Loading fiber onto splicer
Splicing start automatically
Removing spliced fiber
Centering protection sleeve in tube heater
Heating start automatically
Completed
Stripping fiber
Centering spliced point in tube heater
Visual inspection on LCD during splice
Splice Operation
* Hold the fiber at edge of splicer body.
** Apply tension to the fiber until heater
holds the fiber.
• When splicing only standard SM fibers (ITU-T G.652),
“
SM AUTO
” mode is recommended.
• When splicing different types of fibers, “
AUTO
” mode is
recommended, but splice speed is slow.
• When splicer indicates following message, splice
loss may be high
“Arc discharge is not stable. Electrodes should be
stabilized to reform the arc discharge.”
• Load prepared SM fibers onto splicer in order to
stabilize electrodes. After completeing it, re-splice
fibers.
• When an altitude changes drastically, stabilizing
elctrodes must be excecuted before splicing.
Note
Placing Fiber Holder
• Do not reverse Fiber Holder
when placing onto splicer.
Set in the direction which sees
from the sheet key side and
can read the name of a holder.
*Hold
Marking
**Pull
When using single holder
Adjusting the position of
the fiber coating edge to
holder edge
• Make sure the stripped fiber is free
of coating debris or contamination.
• Use only 99% or better purity alcohol.
• Hold lid firmly
• Free of any gel residue
• Hold this point firmly
Fusion Splicer 19R+/70R+