24
TN5A1191-E
3.5 A/D conversion value
Factory mode > 11. A/D data
No Count No Count No Count
0
Ch1 input
value
8
Temperature
sensor 4
16 Ground
1
Ch2 input
value
9 Temperature
sensor 5
17
Resistance
1-2
2
Ch3 input
value
10
Resistance
1-1
18
Resistance
2-2
3
Ch4 input
value
11
Resistance
2-1
19
Resistance
3-2
4
Ch5 input
value
12
Resistance
3-1
20
Resistance
4-2
5 Temperature
sensor 1
13
Resistance
4-1
6 Temperature
sensor 2
14 Pressure
7
Temperature
sensor 3
15
Reference
voltage
Factory mode initial screen
The cursor is in 11.
ENT
ESC
Check that the A/D values during dry zero gas flow fall under the following range.
No. 0 (Infrared ray component 1)
35,000 to 80,000 (ZPA) or 350,000 to 800,000 (ZPB & ZPG)
No. 1 (Infrared ray component 2)
35,000 to 80,000 (ZPA) or 350,000 to 800,000 (ZPB & ZPG)
No. 2 (Infrared ray component 3)
35,000 to 80,000 (ZPA) or 350,000 to 800,000 (ZPB & ZPG)
No. 3 (Infrared ray component 4)
35,000 to 80,000 (ZPA) or 350,000 to 800,000 (ZPB & ZPG)
No. 4 (O
2
sensor input)
18,000 to 22,000
Ignore the values for the channels (components) you do not use.