Problem
Solution
Unable to create new
exams
l
Ensure that a transducer is connected to the front
panel of the Vevo MD Imaging System, and ensure
that it has been initialized.
Unable to archive an
exam
l
Ensure that the Ethernet cable is attached and that
the connection is active.
Patient management troubleshooting
Problem
Solution
Lack of penetration or
sensitivity
l
Ensure that there is adequate coupling medium (for
example, ultrasound gel) between the transducer
and the patient.
l
Adjust the position of the TGC sliders.
l
Ensure the appropriate transducer is being used.
2D and M-Mode troubleshooting
Problem
Solution
Color bands in the
image
l
Adjust wall filter setting.
l
Adjust the priority setting.
l
Adjust line density setting.
Respiration artifacts in
the image
l
Adjust wall filter setting.
Color Doppler Mode troubleshooting
412
Appendices