Isolating the Vevo MD Imaging System from power
The Vevo MD Imaging System does not become completely isolated from power by
pressing the power off button. The only way to completely remove power is to perform
the following procedure.
To isolate the system from power
1
Press the power button.
2
Wait for the system to power down.
3
Set the main power switch on the system base to the off position. See
4
Disconnect the AC input power cord from the stand base.