Cleaning and disinfecting
8-3
Clean and disinfect system and transducer to a high level
(semi-critical uses)
Use this procedure to clean and high-level disinfect the ultrasound system and transducer
whenever it
has
come into contact with blood, broken skin, mucosal membranes, or bodily fluids
.
Follow the manufacturer's instructions when using cleaners and disinfectants. The cleaners and disinfectants
listed in the procedure are both chemically compatible and have been tested for efficacy with the system
and transducers. Confirm that the cleaners and disinfectants are appropriate for your facility’s use.
To clean and disinfect the system and transducer
1 Turn off
the system by pressing the
Power button.
2 Unplug
the power cord from the outlet.
3 Remove
the disposable transducer sheath, if applicable.
WARNINGS
To avoid electrical shock, before cleaning, turn off the system and disconnect it
from the power supply.
Wear the appropriate personal protective equipment (PPE) recommended by
the chemical manufacturer, such as eyewear and gloves.
Cautions
Do not skip any steps or abbreviate the cleaning and disinfecting process in any
way.
Do not spray cleaners or disinfectants directly on the system surfaces or on
system and transducer connectors. Doing so may cause solution to leak into the
system, damaging it and voiding the warranty.
Do not attempt to disinfect a transducer or transducer cable using a method or
chemical that is not included here. This can damage the transducer and void the
warranty.
Use only FUJIFILM SonoSite recommended cleaners and disinfectants. Using a
non- recommended disinfecting solution or incorrect solution strength can
damage the System and transducer and void the warranty. Follow the
disinfectant manufacturer’s recommendations for solutions strengths.
Notes
You must clean and disinfect both the ultrasound system and the transducer
after each use, but only the transducer can be disinfected to a high level.
To clean the TEExp transducer, refer to the
TEExp Transducer User Guide.
Option A
Summary of Contents for SONOSITE X-PORTE
Page 1: ...USER GUIDE ...
Page 75: ...Setting up the system 3 39 ...
Page 76: ...3 40 Setting up the system ...
Page 128: ...4 52 Imaging ...
Page 186: ...5 58 Measurements and calculations ...
Page 210: ...6 24 Measurement references ...
Page 234: ...8 16 Cleaning and disinfecting ...
Page 262: ...9 28 Safety ...
Page 334: ...10 72 Acoustic output ...
Page 364: ...B 14 Index printing 5 56 zoom 4 23 ...
Page 365: ......
Page 366: ...P14645 05 P14645 05 ...