7-2
Troubleshooting and maintenance
Printing does not work.
If you are printing to a DICOM printer, check with your system administrator to
make sure that DICOM settings are accurate and that your system is connected to the network.
If you are printing to a local printer, do any of the following:
Check the printer connections. The system will automatically detect the printer.
Ensure that the local printer is turned on and set up properly. See the printer manufacturer’s
instructions, if necessary.
Only saved images and video clips and the current patient’s worksheets can be printed.
DVR does not record.
Make sure that the USB memory stick has enough available space.
System does not recognize the transducer.
Disconnect and reconnect the transducer.
Maintenance icon (
) appears on screen.
Restart the system. If the issue recurs, system
maintenance may be required. Note the number that appears in parentheses on the C: line and contact
FUJIFILM SonoSite or your FUJIFILM SonoSite representative.
System prompts you to ensure the USB device is valid.
Use the USB memory stick included with the
system.
Make sure that the USB memory stick is not software-encrypted and is not defective.
System prompts you to ensure the USB device contains valid data.
Make sure that the data are
present on the USB memory stick.
Re-export the original data onto the USB memory stick. Contact your system administrator.
USB device does not appear in the list.
Check that the USB device has been properly inserted into an
available USB slot. Use the USB memory stick that came with the system.
System displays the alert “…internal storage device is full.”
Free internal storage space by ending
the current exam, archiving or exporting patient exams, and then deleting them from the system.
Cannot access Patient Form. Cannot access Patient List.
Ensure that you are logged in as a user, not
as a guest.
System does not export or transfer video clips (DICOM).
In the
Archiver
settings of DICOM Config,
make sure that
Exclude Video Clips
is not checked.
Software licensing
FUJIFILM SonoSite software is controlled by a license key. After you install new software, the system
prompts you for a license key. You must obtain one key for each system and transducer package that uses
the software.
Note
If you are using the AC printer, the system must be connected to AC mains.
Summary of Contents for SONOSITE X-PORTE
Page 1: ...USER GUIDE ...
Page 75: ...Setting up the system 3 39 ...
Page 76: ...3 40 Setting up the system ...
Page 128: ...4 52 Imaging ...
Page 186: ...5 58 Measurements and calculations ...
Page 210: ...6 24 Measurement references ...
Page 234: ...8 16 Cleaning and disinfecting ...
Page 262: ...9 28 Safety ...
Page 334: ...10 72 Acoustic output ...
Page 364: ...B 14 Index printing 5 56 zoom 4 23 ...
Page 365: ......
Page 366: ...P14645 05 P14645 05 ...