Measurements and calculations
5-35
To calculate Tricuspid Annular Plane Systolic Excursion (TAPSE)
TAPSE is used to determine the state of the right ventricular systolic function and to correlate this
measurement with the right ventricular ejection fraction.
1
On a frozen M Mode sweep, tap
Calcs
.
If this control is hidden, tapping
More Controls
displays it.
2
On the calculation list, tap
TAPSE
.
3
On the measurements list, tap
TAPSE
.
4
Position the calipers by dragging.
5
To save the measurements, tap
under the measurement name.
To measure a Tissue Doppler Imaging (TDI) waveform
1
Ensure that TDI is on (see
2
On a frozen Doppler spectral trace, tap
Calcs
.
If this control is hidden, tapping
More Controls
displays it.
3
Do the following for each measurement you want to take:
a
On the TDI calculation list, tap the measurement name.
b
Position the calipers by dragging.
c
Save the calculation: Tap
under the measurement name.
Gynecological calculations
The Gynecological calculations include measurements for uterus, ovaries, and follicles.
In the Gynecological exam, you can also calculate volume (see
WARNINGS
To avoid incorrect calculations, verify that the patient information, date, and
time settings are accurate.
To avoid misdiagnosis or harming the patient outcome, start a new patient
form before starting a new patient exam and performing calculations. Starting
a new patient form clears the previous patient’s data. The previous patient’s
data will be combined with the current patient if the form is not first cleared.
Summary of Contents for SONOSITE X-PORTE
Page 1: ...USER GUIDE ...
Page 75: ...Setting up the system 3 39 ...
Page 76: ...3 40 Setting up the system ...
Page 128: ...4 52 Imaging ...
Page 186: ...5 58 Measurements and calculations ...
Page 210: ...6 24 Measurement references ...
Page 234: ...8 16 Cleaning and disinfecting ...
Page 262: ...9 28 Safety ...
Page 334: ...10 72 Acoustic output ...
Page 364: ...B 14 Index printing 5 56 zoom 4 23 ...
Page 365: ......
Page 366: ...P14645 05 P14645 05 ...