Acoustic output
65
English
De
ut
sch
Español
Fr
a
n
ç
a
is
Ita
lia
no
P
o
rtuguês
Nederlands
Table 27: Transducer model:
rP19x Orbital Use
Operating mode:
PW Doppler
Index label
M.I.
TIS
TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
≤1
A
aprt
>1
Global maximum index value
0.18
—
—
0.27
0.59
0.57
As
so
ciat
ed
aco
u
st
ic
parameter
p
r.3
(MPa)
0.27
W
0
(mW)
—
—
35.3
37.4
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
25.3
z
1
(cm)
2.5
Z
bp
(cm)
2.5
Z
sp
(cm)
3.35
z@PII
.3max
(cm)
3.5
d
eq
(Z
sp
)
(cm)
0.80
F
c
(MHz)
2.23
—
—
2.23
2.23
2.23
Dim of A
aprt
X (cm)
—
—
1.86
1.80
1.86
Y (cm)
—
—
1.15
1.15
1.15
O
ther information
PD
(μsec)
6.557
PRF
(Hz)
1953
p
r
@PII
max
(MPa)
0.36
d
eq
@Pll
max
(cm)
0.64
Focal Length
FL
x
(cm)
—
—
29.82
29.82
FL
y
(cm)
—
—
9.00
9.00
I
PA.3
@MI
max
(W/cm
2
) 2.49
Operating
contr
o
l
co
nditio
ns
Control 1: Exam type
Orb
Orb
Orb
Orb
Control 2: Sample volume size
5 mm
14 mm
14 mm
14 mm
Control 3: Sample volume position Zone 6
Zone 7
Zone 5
Zone 7
Control 4: PRF
1953
1953
1953
1953
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
#
No data are reported for this operating condition since the global maximum index value is not reported for the
reason listed. (Reference Global Maximum Index Value
line.)
— Data are not applicable for this transducer/mode.