![FujiFilm SonoSite SII User Manual Supplement Download Page 65](http://html1.mh-extra.com/html/fujifilm/sonosite-sii/sonosite-sii_user-manual-supplement_2343393065.webp)
En
glish
De
utsch
Español
Fr
a
n
ça
is
It
a
lia
n
o
Po
rt
u
g
u
ê
s
Acoustic output
63
English
De
ut
sch
Español
Fr
a
n
ç
a
is
Ita
lia
no
P
o
rtuguês
Nederlands
Table 25: Transducer Model:
P10x
Operating Mode:
CW Doppler
Index label
M.I.
TIS
TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
≤1
A
aprt
>1
Global maximum index value
(a)
—
(a)
—
1.8
1.7
As
so
ciat
ed
aco
u
st
ic
parameter
p
r.3
(MPa)
2.59
W
0
(mW)
—
#
34.8
25.7
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
—
z
1
(cm)
—
Z
bp
(cm)
—
Z
sp
(cm)
0.70
z@PII
.3max
(cm)
#
d
eq
(Z
sp
)
(cm)
0.36
F
c
(MHz)
#
—
#
—
4.00
4.00
Dim of A
aprt
X (cm)
—
#
—
0.32
0.16
Y (cm)
—
#
—
0.70
0.70
O
ther information
PD
(μsec)
#
PRF
(Hz)
#
p
r
@PII
max
(MPa)
#
d
eq
@Pll
max
(cm)
0.27
Focal Length
FL
x
(cm)
—
#
—
0.92
FL
y
(cm)
—
#
—
5.00
I
PA.3
@MI
max
(W/cm
2
) #
Operating
contr
o
l
co
nditio
ns Control 1: Exam type
Crd
Crd
Control 2: Sample volume position
Zone 0
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
#
No data are reported for this operating condition since the global maximum index value is not reported for the
reason listed. (Reference Global Maximum Index Value
line.)
— Data are not applicable for this transducer/mode.
Summary of Contents for SonoSite SII
Page 1: ...Doppler and ECG User Guide Supplement...
Page 136: ...134 Schall Ausgangsleistung...
Page 270: ...268 Puissance acoustique...
Page 338: ...336 Uscita acustica...
Page 406: ...404 Sa da ac stica...
Page 475: ......
Page 476: ...P21465 05 P21465 05...