Acoustic output
61
English
De
ut
sch
Español
Fr
a
n
ç
a
is
Ita
lia
no
P
o
rtuguês
Nederlands
Table 23: Transducer model:
L38xi
Operating mode:
PW Doppler
Index label
M.I.
TIS
TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
≤1
A
aprt
>1
Global maximum index value
1.3
—
2.6
—
3.7
(b)
As
so
ciat
ed
aco
u
st
ic
parameter
p
r.3
(MPa)
2.59
W
0
(mW)
—
114.5
114.5
#
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
—
z
1
(cm)
—
Z
bp
(cm)
—
Z
sp
(cm)
1.20
z@PII
.3max
(cm)
0.7
d
eq
(Z
sp
)
(cm)
0.32
F
c
(MHz)
4.06
—
4.78
—
4.78
#
Dim of A
aprt
X (cm)
—
1.86
—
1.86
#
Y (cm)
—
0.40
—
0.40
#
O
ther information
PD
(μsec)
1.230
PRF
(Hz)
1008
p
r
@PII
max
(MPa)
2.86
d
eq
@Pll
max
(cm)
0.46
Focal Length
FL
x
(cm)
—
5.54
—
#
FL
y
(cm)
—
1.50
—
#
I
PA.3
@MI
max
(W/cm
2
) 323
Operating
contr
o
l
co
nditio
ns
Control 1: Exam type
Art
Nrv
Nrv
Control 2: Sample volume size
1 mm
1 mm
1 mm
Control 3: Sample volume position Zone 0
Zone 7
Zone 7
Control 4: PRF
1008
10417
10417
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
#
No data are reported for this operating condition since the global maximum index value is not reported for the
reason listed. (Reference Global Maximum Index Value
line.)
— Data are not applicable for this transducer/mode.