4-6
4 Applicable usage methods
897N201058J FDC NX700 Instruction Manual 2022.04
4.1.2 Configuring calibrator concentration
Correct the calibration concentration when the concentration included with the calibrator CP to be
used and the concentration in the calibrator concentration display area at the Calibration Lot No.
screen are different.
1. Tap the [Settings] button.
The administrator password entry screen ap-
pears.
2. Enter the administrator password, and
then tap the [OK] button.
The Calibration concentration configuration
screen appears.
NOTE
The administrator password is neces-
sary when moving from the Function
menu screen to the Administrator function
screen.
4