9-3
9 Function menu
897N201058J FDC NX700 Instruction Manual 2022.04
9.1 Control
This mode is used for measurement of control serum (FUJI DRI-CHEM CONTROL QP) and for per-
forming control surveys. When control mode is set, measurements are performed under the condi-
tions listed below.
•
Correlation coefficients: follow [Control measurement setting] (Initial values disabled (a=1, b=0))
of the function menu.
• Measurement units: follow [Unit settings] of the function menu.
•
Dilution conditions: dilution is not normally performed. (This can be configured at the Sample
information registration screen.)
NOTE
When the analyzer is switched on again, control mode is canceled.
1. Prepare the slide to be measured, and
then load it in the sample position to be
measured.
2. Place the control liquid to be measured
in the tube, and then load it in the sam-
ple position to be measured.
3. Tap the button.
4. Tap the [Control] button.
9