4-27
897N203029A FDC NX600 Instruction Manual 2022.01
4 Applicable Usage Methods
4.5 Slide Loading Order When Measuring Different Items Simul
-
taneously
NOTE
Any order of slide stacking can be measured. However, please follow the recommended
slide stacking rules below. Otherwise, the analyzer may consume excessive amounts of dis-
posal tips and cups.
(1) Measuring CRP and CM (colorimetric) slides
CRP
一般比色
図-4.5.2-1
CRP
CM slide
Always put a CRP slide first or last.
If a CRP slide is loaded between CM slides, an ex-
tra tip will be used.
(2) Measuring CM and ISE slides
電解質
一般比色
図-4.5.2-1
ISE
CM slide
We recommend loading an ISE slide last.
If an ISE slide is measured first, it might take extra
time for measurement.
If a CM slide is loaded after an ISE slide, an extra
tip will be used.
(3) Measuring CRP, CM, and ISE slides
CRP
電解質
一般比色
図-4.5.2-4
ISE
CRP
CM slide
Stack CM slides before CRP and ISE slides.
(4) Measuring CM slides that require dilution
5倍希釈
2倍希釈
図-4.5.2-2
5x dilution
2x dilution
If slides are configured with different dilution rates,
stack the slides so that slides with same dilution
factor are measured successively. Otherwise, extra
tips and mixing cups will be used.
NOTE
If all CM slides have the same dilution fac-
tor, there are no restrictions on the stacking
order.
<Not recommended slide loading order>
CRP
電解質
一般比色
図-4.5.2-4
CRP
CM (Dil)
CM slide
Only one loading position is prepared for the diluent
in this analyzer. In the example, diluent (purified
water etc.) for the colorimetric slide that requires di-
lution [CM (Dil)] is loaded and the measurement will
start. The next measurement requires the specified
diluent for CRP, so that an error (E0160 Incorrect
diluent type error) occurs before the measurement,
and the measurement will be completed. Rerun the
CRP measurement from the beginning.
4