4 Applicable Usage Methods
4.2.2 Using a Pipette
WARNING:
When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab
coat, and safety goggles) in accordance with the sample handling rules of your facility. If you acci-
dentally touch a sample, immediately rinse the contaminated area thoroughly under running water,
and then disinfect the area. Seek medical assistance if necessary.
IMPORTANT:
Mount the tip securely into the nozzle of the pipette.
If mounting is poor, the tip may be dislodged, and the sample may scatter.
Applying samples to slides
液滴を作ってスライドに近づけます
液滴をスライドの中央に点着します
スライドが検体を吸い取りながら
内部に拡散していきます
0057
Make a drop and then move it near the slide
1. Move the tip of the pipette near the slide.
2. Slowly push the push button and make a
drop.
液滴を作ってスライドに近づけます
液滴をスライドの中央に点着します
スライドが検体を吸い取りながら
内部に拡散していきます
0057
Place the drop on the center of the slide
液滴を作ってスライドに近づけます
液滴をスライドの中央に点着します
スライドが検体を吸い取りながら
内部に拡散していきます
0057
While the slide is absorbing the
sample, it diffuses inside
3. Place the drop on the center of the slide.
IMPORTANT:
Always make a drop at the top of the tip, and then
apply the sample to the slide. If the sample is
discharged to the slide surface without making a
drop, the measurement results may be affected.