3 Measurement
3.5 ISE Tests
IMPORTANT:
Analyzer cannot detect expiries of ISE slides if the installed software is version 1.0. Be sure manu-
ally that slides are still valid.
Otherwise, with the software version 1.1 or later is installed, analyzer can automatically detect ex-
piries of the ISE slides that will expire 2023 or later or whose lot numbers are 130000 to 139999.
NOTE
When urine (U) is selected as the sample type for measurement, the dilution setting is set to
2x automatically. Therefore, diluent is needed in addition to reference fluid. The data is cal
-
culated automatically, so the results can be reported without adjustment.
NOTE
Refer to the “Instructions for Use” of the slide for usable diluent for urine measurements.
NOTE
Two tips are necessary for each slide; one for sample, the other for reference fluid.
Reference fluid
1. Load the reference fluid.
For the reference fluid container (FUJI plain
tube 0.5 mL or 1.5 mL), use the container con-
figured for [Tube setting for diluent & reference
fluid] in Mode function for administrator (
Section
9.3.15).
For details of the loading method, refer to Sec-
2. Check quantity of tips.
• Loading tips
3. Load diluent for urine sample measure-
ment.
Also, check the quantity of mixing cups. If the
number is insufficient, replace with new ones.
• Loading mixing cups
• Loading diluent
NOTE
When performing an ISE measurement
that requires dilution, use at least 3 tips
and 1 well.