2-5
897N203029A FDC NX600 Instruction Manual 2022.01
2 Part Names and Consumables
2.2 Touch Panel Screen Names and Functions
2.2.1 Measurement Preparation Screen (Top Screen)
If the power is switched on, the Measurement-preparation screen (top screen) is displayed after startup.
1
8
a
0
9
2
3
6
7
4
5
The touch panel does not have multi-touch input
function.
Do not touch plural points on the panel at the
same time.
1
Help button
Display the help screen.
2
Function menu button
Display the function menu screen.
Refer to Chapter 9
3
Work list button
Obtain order information from the host PC
via a DMS connection.
4
Clear button
Clear the entered information.
5
Plasma filter (PF) production num-
ber (lot) display
Displays the lot number of the plasma filter
(PF) in use. Make sure this number matches
the production number (lot) of the plasma
filter (PF) in use.
6
Mixing cups
Displays the quantity of mixing cups. Tap to
replace the mixing cups.
7
Tips
Tap to replenish tips.
8
[START] button
Start a measurement.
9
Operator ID button (displayed ac-
cording to settings)
If tapped, you can enter the operator ID.
0
Feed button
Feed the recording paper a fixed length.
[STOP] button (displayed during
sample application operation)
If the [STOP] button is tapped during sample
application, tips are discarded, operation is
stopped, and the device enters a paused state.
a
Measurement results button
Display the measurement results.
CAUTION:
2