![FujiFilm FUJI DRI-CHEM NX600 Instruction Manual Download Page 200](http://html1.mh-extra.com/html/fujifilm/fuji-dri-chem-nx600/fuji-dri-chem-nx600_instruction-manual_2343341200.webp)
8 Troubleshooting
8-36
897N203029A FDC NX600 Instruction Manual 2022.01
8.2.12 PF Errors
Error
code
Error message
Error condition
Troubleshooting
E0132
PF pressure ERR
Replace the PF pump
tube.
A pressure error oc-
curred during the filter
-
ing process.
Replace the PF Tube.
E0133,
E0150
PF sensor ERR
Contact your dealer.
A PF sensor error has
occurred.
Please contact your dealer for assis-
tance.
W0134
PF sampling ERR
Error occurred during
PF sampling.
Check for a PF leak.
An error occurred dur-
ing PF filtering.
Clean the PF Suction Pad.
W0138
U n a c c e p t a b l e P F
sampling WARNING
For PF measurement,
turn the power SW
OFF and then ON.
Contact your dealer in
case the ERR repeats.
Measurements other
than PF measurements
can be measured with-
out turning the power
SW OFF and then ON.
A PF measurement
error has occurred,
and PF measurement
cannot be started.
Check PF measurement errors that oc-
curred before this error, and then carry
out the corresponding troubleshooting.
After troubleshooting, turn the analyzer
on and off, and then start measurement
again from the beginning.
E1300,
E1301,
E1303,
E1304
PF motor ERR
Problem with the plas-
ma filter.
Contact your dealer.
PF motor error move-
ment error
Turn the power on and off. If the error
occurs again, please contact your dealer
for assistance.
W1308
PF sensor ERR
Problem with the plas-
ma filter.
Contact your dealer.
A PF sensor error has
occurred.
Please contact your dealer for assis-
tance.
E1302
PF motor operation
ERR
Problem with the plas-
ma filter.
Contact your dealer.
PF motor error move-
ment error
Turn the power on and off. If the error
occurs again, please contact your dealer
for assistance.
8