8 Troubleshooting
8-32
897N203029A FDC NX600 Instruction Manual 2022.01
Error
code
Error message
Error condition
Troubleshooting
E0160
Incorrect diluent type
error
Diluent type is incor-
rect.
Use specified diluent
for the test name.
Diluent type does not
match the slide to be
used.
Use specified diluent for the test name,
and rerun the measurement from the
beginning.
W0175
No slide
Slides have not been
set.
Rerun the tests from
the beginning.
No slides are loaded in
the slide cartridge.
Load slides to be tested, and rerun the
tests from the beginning.
W5400
Expired slide
Slide is expired and
cannot be used.
The validity term of the
slide has expired.
Always use valid slides.
IMPORTANT:
Slide validity may not be detected in
ISE tests.
NOTE
Depending on the device set-
tings, this warning may not be
generated.
8