8 Troubleshooting
8-31
897N203029A FDC NX600 Instruction Manual 2022.01
8.2.10 Slide Loading Errors
Error
code
Error message
Error condition
Troubleshooting
W0040
I n c o r r e c t s a m p l e
WARNING
The sample type set
in analyzer does not
correspond to the read
sample type.
Set the sample type
by SAMPLE key and
the slide’s sample type
corresponding to the
sample. Rerun the
tests from the begin-
ning.
The sample type se-
lected by the [Sample
type] button and the
sample type of the
slide do not match.
(Example)
• Whole blood (W)
slides were loaded
when either [P/S] or
[U] had been se-
lected by the [Sample
type] button.
• Plasma/serum (P)
slides were loaded
when [W] had been
selected by the
[Sample type] button.
Match the configuration by the [Sample
type] button and the sample type to
measure the slide to be used, and then
rerun the measurement from the begin-
ning.
W0041
Incorrect sample type
WARNING
Although a PF sam-
ple rack is loaded, the
sample type setting is
not P/S.
The sample type set-
ting of PF measure-
ment does not match.
Set the sample type to P/S.
W0042
Incorrect sample rack
WARNING
Correct sample rack
not set in place.
Check the sample
rack.
A re-run test for a
measurement by us-
ing PF is attempted,
but a rack other than
a PF sample rack was
loaded.
Load a PF sample rack in which a
sample extracted by PF has been
loaded.
A PF sample rack has
been loaded for a con-
trol measurement.
Load a sample rack other than a PF
sample rack.
A sample rack other
than for CP has been
loaded for a calibration
measurement.
Load a sample rack for calibrator CP.
W0043
N o s a m p l e r a c k
WARNING
Sample rack is not set.
Set a sample rack.
No sample rack is
loaded.
Set a sample rack.
W0090
ISE slide direction
ERR
Incorrect loading direc-
tion of ISE slide.
Check the ISE slide.
The ISE slide is loaded
in reverse.
Load the ISE slide in the slide cartridge
in the proper direction, and then meas-
ure again.
8