8 Troubleshooting
8-25
897N203029A FDC NX600 Instruction Manual 2022.01
Error
code
Error message
Error condition
Troubleshooting
E0540
ISE AD conversion
initialization ERR
For ISE measurement,
turn the power SW
OFF and ON.
Contact your dealer in
case the ERR repeats.
Other parameters
besides ISE can be
measured.
A circuit board signal
error has occurred.
When an error occurs after turning the
power on and off, please contact your
dealer for assistance. Tests other than
ISE do not require a restart.
E1530,
E1531,
E1532
Temperature AD con-
version ERR
Contact your dealer.
Please contact your dealer for assis-
tance.
E1750-
E1755,
E1757
Slide reading unit
failure
Contact your dealer.
If the error still occurs after switching
the analyzer off and on, please contact
your dealer for assistance.
E3900
Internal communica-
tion ERR
Contact your dealer.
E4401,
E4403
Save ERR
Contact your dealer.
E4410
Measurement result
reference error
The number of meas-
urement results, with
the new result, current-
ly exceeds capacity.
The selected measure-
ment result was de-
leted.
This measurement result cannot be
referred.
E4411
Control chart refer-
ence error
The selected chart is
not up-to-date with the
new measurement.
Please go back to the
parameter selection
screen and display the
chart again.
Please go back to the parameter selec-
tion screen and display the chart again.
8