8 Troubleshooting
8-24
897N203029A FDC NX600 Instruction Manual 2022.01
8.2.8 Circuit Board Signal Errors
Error
code
Error message
Error condition
Troubleshooting
E0035
Light too intense
Light too intense.
Contact your dealer in
case the ERR repeats.
A circuit board signal
error has occurred.
If the error still occurs after switching
the analyzer off and on, please contact
your dealer for assistance.
E0300,
E0301,
E0302
Gain setting ERR
Contact your dealer in
case the ERR repeats.
E0509
AD conversion cali-
bration ERR
Contact your dealer.
Please contact your dealer for assis-
tance.
E0510
Faulty A/D controller
Contact your dealer.
E0515
DA conversion ini-
tialization ERR
Contact your dealer.
E0516
AD conversion reset
ERR
Contact your dealer.
E0517
AD conversion ini-
tialization ERR
Contact your dealer.
E0532,
E0533,
E0537,
E0538
ISE self test AD con-
version ERR
For ISE measurement,
turn the power SW
OFF and ON.
Contact your dealer in
case the ERR repeats.
Other parameters
besides ISE can be
measured.
When an error occurs after turning the
power on and off, please contact your
dealer for assistance. Tests other than
ISE do not require a restart.
E0534,
E0535,
E0536,
E0539
ISE AD conversion
ERR
For ISE measurement,
turn the power SW
OFF and ON.
Contact your dealer in
case the ERR repeats.
Other parameters
besides ISE can be
measured.
8