8 Troubleshooting
8-18
897N203029A FDC NX600 Instruction Manual 2022.01
(9) During a sampler drive system error
Error
code
Error message
Error condition
Troubleshooting
W0070
Spotting position
ERR
Sampler rotation error.
Rotation error of the
sampler has occurred.
Turn the analyzer off after the meas
-
urement, and make sure that foreign
matters such as tips and slides are not
stuck or left inside of the sampler.
NOTE
If the error still occurs after
switching the analyzer off and
on, please contact your dealer
for assistance.
E1000,
E1001,
E1003,
W1004
Sampler up/down
motor ERR
Problem moving the
sampler up and down.
Perform inspection.
Contact your dealer.
Up-and-down move-
ment error of the motor
E1002
Sampler up/down
motor operation ERR
Problem moving the
sampler up and down.
Perform inspection.
Contact your dealer.
E1100-
E1103,
W1104
Sampler movement
motor ERR
Problem with sampler
movement.
Perform inspection.
Contact your dealer.
Rotation error of the
sampler
E1200,
E1201,
E1203,
W1204
Syringe motor ERR
Problem with the sy-
ringe.
Contact your dealer.
Movement error of the
syringe motor
E1202
Syringe motor opera-
tion ERR
Problem with the sy-
ringe.
Contact your dealer.
8