1-9
897N203029A FDC NX600 Instruction Manual 2022.01
1 Safe Usage and Handling Precautions
Do not touch used tips, tubes, and mixing cups with your bare hands as this may cause infection. If
you accidentally touch an item, immediately rinse the contaminated area thoroughly under running
water, and then disinfect the area. Seek medical assistance if necessary.
Used tips, tubes, and mixing cups are infectious industrial waste. Process this waste in compliance
with any applicable regulations in your country, such as by incineration, melting, sterilization, or dis-
infection. Furthermore, when entrusting disposal to another party, request processing from a con-
tractor with a license for special management of industrial waste processing work.
Always use new FUJI DRI-CHEM AUTO TIPS, FUJI Sample Tubes, blood collection tubes, and
FUJI DRI-CHEM MIXING CUPS for each sample. Do not reuse old ones.
Use only specified products. Otherwise, errors or malfunctions may occur.
When the power is shut off during test processing, such as by a power outage, replenish tips and
replace mixing cups. The analyzer may lose the information of the usage positions of the consuma-
bles and use used mixing cups (holes).
1.19 Light Source Lamp
The light source lamp gets very hot. Before replacing the lamp, turn the power off, wait at least 5
minutes, and then check that the lamp has cooled.
(1) The light source lamp is a precision halogen lamp. Do not touch the glass surface of the lamp
with your bare hands.
(2) The lamp is a consumable with a lifetime. Always keep a spare lamp on hand in case the lamp
burns out.
1.20 Recording Paper
(1) Use specified recording paper for the FDC NX600.
(2) If other recording paper is used, the printer head section may be damaged.
WARNING:
WARNING:
IMPORTANT:
IMPORTANT:
IMPORTANT:
CAUTION: