1-6
897N203029A FDC NX600 Instruction Manual 2022.01
1 Safe Usage and Handling Precautions
1.10 Measurement Results
Make a diagnosis in a comprehensive manner, considering other relevant test results or clinical symptoms.
When an equipment problem (error) appears before or during measurement, or a warning appears
in the measurement results, the measurement results may not be accurate. Refer to Chapter 8,
Troubleshooting, and then rerun the measurement.
To maintain the accuracy of your test results, it is recommended to perform quality control on routine
bases by use of control fluids. (Refer to
Chapter 7.)
1.11 FUJI DRI-CHEM Slides
NOTE
For the remainder of the manual, FUJI DRI-CHEM slide is shortened to “slide”.
Please confirm about the usable slide items in this device to your dealer for assistance.
Do not touch used slides with bare hands as this may cause infection. If you accidentally touch an
item, immediately rinse the contaminated area thoroughly under running water, and then disinfect
the area. Seek medical assistance if necessary.
Used slides are infectious industrial waste. Process this waste in compliance with any applicable
regulations in your country, such as by incineration, melting, sterilization, or disinfection. Further-
more, when entrusting disposal to another party, request processing from a contractor with a license
for special management of industrial waste processing work.
Please read the Instructions for Use of the slides carefully and follow the instructions.
• The slides should be stored in a refrigerator still wrapped to avoid humidity, light, and heat.
• Take out the required number of slides from the refrigerator, and allow them to reach room tem-
perature for 5 minutes or more with the condition of the individual package before unwrapping and
using them. Otherwise, the measurement results and the analyzer operation may be affected.
• Complete the measurement within 30 minutes after unwrapping the individual package.
• Do not touch the center part (both sides) of a colorimetric slide.
• Do not touch the thread bridge part of an electrolyte slide.
•
Do not scratch the identification information on the back of a slide.
• Do not reuse a slide.
NOTE
Slide packages and containers for liquids such as diluent and reference fluid may change
without notice.
IMPORTANT:
IMPORTANT:
IMPORTANT:
CAUTION:
WARNING:
WARNING:
CAUTION:
IMPORTANT:
1