6-9
897N203029A FDC NX600 Instruction Manual 2022.01
6 Maintenance
6.3.2 Cleaning the Transfer Bar, Incubator, Sample Application Unit, and
ISE Unit
(1) Cleaning the transfer bar
1. Tap the [Cleaning mode] button in Normal
mode.
2. Remove the slide cartridge and the tip
rack from the analyzer.
3. Tap the [Start Cleaning] button.
NOTE
Keep your fingers clear as the analyzer
runs.
6