4-28
897N203029A FDC NX600 Instruction Manual 2022.01
4 Applicable Usage Methods
4.6 Automatic Acquisition of Sample Information (Work Lists)
If the analyzer is connected to DMS, you can receive patient IDs and their sample information by
tapping the button at the Measurement-preparation screen before measurement.
NOTE
DMS is sold separately.
1. Tap the button. The analyzer receives
a work list (a list of patients for the sam-
ples to be measured) from DMS.
The Work list index screen appears.
NOTE
If there are no measurement requests,
the buzzer sounds, and the message
[There is no request information] is dis-
played.
NOTE
The display of the list is configured by
[Work list selection setting] in Mode
function for administrator.
NOTE
On DMS, "Sample No." is "Sample ID"
and "Sample ID" is "Patient ID”.
NOTE
When there are multiple pages in a
list, you can scroll the pages using the
/
buttons.
NOTE
If the [Cancel Work list] button is tapped,
the analyzer returns to the Measure-
ment preparation screen.
2. Tap the sample information you want to
register.
4