5.7 Processor Adjustment/Maintenance (06)
5
This procedure calibrates the processor section
temperatures.
[Selection]
[Setup and Maintenance]
→
[06 Processor
Adjustment/Maintenance]
→
[0623 Processor
Temperature Calibration]
1. Check that the display values have reached the
“Setting” temperatures.
2. Open the dryer unit and support it with Stopper 2
(yellow).
3. Open the replenisher box door.
4. Remove the circulation filter tray.
5. Remove the P1 circulation filter and put it on the
circulation filter tray.
6. Measure the temperature of the P1 processing
solution and record it.
7. Reinstall the filter by aligning the fastening tab
with the notch, and then turn it 90° clockwise.
8. If the “Setting” temperatures differ from the values
displayed, move the cursor to the “Measured
value” and enter the values recorded in Step 4.
Range: 0.0 to 99.9
9. Click the [Enter] button.
10. Repeat Steps 4 through 9 for the P2 solution
temperature.
5.7.4
Processor Temperature Calibration (0623)
L
2
233
P1 Circulation Filter
Circulation Filter Tray
L
2
1047
Thermometer
P1 Sub-tank
P2 Sub-tank