5-43
5.4 Scanner Adjustment/Maintenance (03)
5
This procedure displays the current status of each
sensor.
NOTE: Refer to Section 18.1 for input parts location.
[Selection]
[Setup and Maintenance]
→
[03 Scanner Adjustment/
Maintenance]
→
[0320 Input Check]
1. Click the [Next Page] button.
2. Click the [Close] button.
• Operation returns to the menu screen.
This menu is used to set up the center value for the
auto focusing of each film carrier mask.
[Selection]
[Setup and Maintenance]
→
[03 Scanner Adjustment/
Maintenance]
→
[0321 Focus Position Adjustment]
1. Install the carrier and proceed to the “Focus
Position Adjustment” screen.
• Carrier ID, type and mask type (M69D only) are
displayed.
2. Select “135F” or “135H” for the NC100AY.
5.4.2
Input Check (0320)
↓
[Next] [Previous]
↑
5.4.3
Focus Position Adjustment (0321)