INZ-TN7ZSS-E
- 69 -
6.5 Parameter setting
On the “Parameter Setting” screen, “Path Lengths” related to measurement value and “Average Period” re-
lated to average value output are set. Items to be set are as follows.
Explanation of setting items
“Path Lengths” : Enters the optical path lengths.
“Average Period” : Sets how many minutes the value is output as an average value of movement in the
case of the average value output.
“Reset Avg.Gas” : Resets the average value in the case of the average value output.
“Key Lock”
: Any key operation except the key lock OFF cannot be performed.
“Disp.Off”
: Sets automatic OFF of the backlight of display unit and the time until backlight.
“Date/Time”
: Sets the current year/month/date, hour: minute: second.
Select the “Parameter Setting” from the “Menu” screen, and press the
key.
6.5.1 Setting of stack lengths
Enter the lengths of the stack where the receiver unit and the transmitter unit are attached. It does not include
the lengths of companion flange. This value has a direct effect on the measurement value, so be sure to enter
correctly. Otherwise, measurement value will not be properly displayed or output. Enter the value in the mil-
limeter. (Input range is from 100 to 15000mm.)
Explanation of concentration output
The analyzer is calculated and output as a standard of the 1m of stack lengths (optical path lengths).
Absorption concentration varies depending on the stack lengths, so be sure to enter the stack lengths cor-
rectly.
ENT
2006-04-01
12:00:00
mm
01000
HC
Min.
01
%
12
OFF
OFF
09
10/06/02 16:00:19
Min.
2006-04-01
12:00:00
Menu
Zero Calibration
Alarm Log
Display/Output
Pasword
Parameter Setting
Input Signal
Span Calibration
Alarm Setting
AO Select
Output Hold
Analyzer Information
Average Period
cv O
2
Reset Avg.Gas
Key Lock
Disp.OFF
Date/Time
Parameter Setting
Path lengths