-80-
4.13.7. Setting the operation mode (normal/high-speed)
Description
Used to switch computation cycle and output cycle.
Settable range
NORMAL
: Standard mode (factory-set value), computation/output cycle is approximately 0.5 seconds.
HIGH SPEED : High speed response mode, computation/output cycle is approximately 0.2 seconds.
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.5.1.)
Operation
(example)
Switch the operation mode to the high speed response mode.
Key operation
Description
Display
▼
Press the
key twice to display “OUTPUT SETUP”.
OUTPUT SETUP
ENT
▼
Press the
ENT
key once to display “ZERO ADJUSTMENT”.
ZERO ADJUSTMENT
SET ZERO
▼
Press the
key for 14 times to display “OPERATION MODE”.
OPERATION MODE
NORMAL
ENT
▼
Press the
ENT
key once to blink the cursor.
OPERATION MODE
NORMAL
▼
Press the
key for 6 times to move the cursor.
OPERATION MODE
HIGH SPEED
Press the
ENT
key once to register.
OPERATION MODE
COMPLETE
↓
ENT
▼
▼
▼
▼
―――
“OPERATION MODE” has been registered.
―――
OPERATION MODE
HIGH SPEED
ESC
Press the
ESC
key once and then press the
key for 3 times to enter
the measurement mode.
0.000 %
0.000 m3
Reference
The difference between standard mode and high speed mode
High speed mode is unfit for the measurement when foreign objects or air bubbles are contained.
Standard mode is about 10 times more resistant to entry of foreign objects or air bubbles than high speed mode.