-78-
4.13.5.8. Temperature alarm
Description
TEMP.ALARM
: Contact outputs DO1, DO2, DO3, and/or DO4 are activated when: the temperature is outside the measuring
range(-40 to 200
℃
), a break in the resistance bulb occurs, or temperature circuit is abnormal. (This function is used only in 1-
path/energy measurement.
)
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.5.1.)
Operation
(example)
Set DO1 OUT to “TEMP.ALARM” and contact action to “ACTIVE ON”.
Key operation
Description
Display
▼
Press the
key twice to display “OUTPUT SETUP”.
OUTPUT SETUP
ENT
▼
Press the
ENT
key once to display “ZERO ADJUSTMENT”.
ZERO ADJUSTMENT
SET ZERO
▼
Press the
key for 8 times to display “DO1 OUT”.
* Use the
key to select “DO2 OUT”, “DO3 OUT”, or “DO4 OUT”.
DO1 OUT
NOT USE
ENT
▼
Press the
ENT
key once to blink the cursor.
DO1 OUT
NOT USE
▼
Press the
key 18 times to display “TEMP.ALARM” on the second
line.
DO1 OUT
TEMP.ALARM
Press the
ENT
key once to register “TEMP.ALARM”.
TEMP.ALARM
**
COMPLETE
**
↓
ENT
▼
▼
▼
▼
―――
“TEMP.ALARM” has been registered.
―――
STATUS OUT
CONTACT ACTION
ENT
▼
Press the
ENT
key once to display “CONTACT ACTION”.
CONTACT ACTION
ACTIVE ON
Press the
ENT
key once to register “ACTIVE ON (Normal OFF).
* Use the
key when selecting “ACTIVE OFF” (Normal ON).
CONTACT ACTION
**
COMPLETE
**
↓
ENT
▼
▼
▼
▼
―――
“ACTIVE ON” has been registered.
―――
STATUS OUT
CONTACT ACTION
ESC
Press the
ESC
key twice, and the
key 3 times to return to the
measurement mode.
0.000 m/s
0.000 m3