- 62 -
INZ-TN1ZSS-E
Analog Input
GasPress.(kPa)
Fixed
Fixed
4mA
20mA
00.00
00.00
10.00
Gas Temp.(
℃
)
0027
0000
0500
O2
(vol%)
20.00
00.00
21.00
H2O
(vol%)
020.0
000.0
100.0
Gas Flow(m/s)
02.00
00.00
20.00
Air Purge(kPa)
CH2
000.0
100.0
0
05.0
Alarm
4mA
20mA
Under
2007-4-1
12:00:00
CH1
Fixed
Fixed
Fixed
Enter the purge pressure value to output alarm.
Change the numeric value by the
key or
the
key, and move the digits by the
key. Press the
ENT
key to validate the set input
value.
(
)
ENT
Analog Input
GasPress.(kPa)
Fixed
Fixed
4mA
20mA
00.00
00.00
10.00
Gas Temp.(
℃
)
0027
0000
0500
O2
(vol%)
20.00
00.00
21.00
H2O
(vol%)
020.0
000.0
100.0
Gas Flow(m/s)
02.00
00.00
20.00
Air Purge(kPa)
CH2
0
00.0
100.0
005.0
Alarm
4mA
20mA
Under
2007-4-1
12:00:00
CH1
Fixed
Fixed
Fixed
Enter the pressure value corresponding to 4,
20mA DC output in the same manner.
6.8 Selecting analog output
There are two types of analog output (4 to 20mA DC output) as a standard specification. Measurement value
and O
2
conversion value can be output individually.
Default value is set to “No”, so be sure to set it after installation.
Select “Analog Output” from the “Menu” screen, and press the
ENT
key.
2006-04-01
12:00:00
Menu
Zero Calibration
Alarm Log
Display
/
Output
Password
*****
Parameter Setting
Span Calibration
Alarm Setting
Output Hold
AO Select
Input Signal
ENT
Analog Output
Analog Output
1(AO1)
HC
ℓ
Analog Output
2 (AO2)
2007-
4
-1
12:00:00
cv
HC
ℓ