- 60 -
INZ-TN1ZSS-E
6.7.2 Setting of analog input (fixed value) (in case of H
2
O)
Sets the fixed value to perform the calibration when the value for which the measured gas condition do not
change, or the sensors to calibrate are not provided, or the number of calibration input terminals is insuffi-
cient. When the actual value differs greatly from the setting value, analog input cannot be measured cor-
rectly.
Analog Input
GasPress.(kPa)
Fixed
Fixed
4mA
20mA
00.00
00.00
10.00
Gas Temp.(
℃
)
0027
0000
0500
O2
(vol%)
20.00
00.00
21.00
H2O
(vol%)
050.0
000.0
100.0
Gas Flow(m/s)
02.00
00.00
20.00
Air Purge(kPa)
No
000.0
100.0
005.0
Alarm
4mA
20mA
Under
2007-4-1
12:00:00
CH1
Fixed
Fixed
CH2
Select “H
2
O” by the
key or the
key.
Press the
key to move the cursor to
“Channel” setting.
(
)
Analog Input
GasPress.(kPa)
Fixed
Fixed
4mA
20mA
00.00
00.00
10.00
Gas Temp.(
℃
)
0027
0000
0500
O2
(vol%)
20.00
00.00
21.00
H2O
(vol%)
050.0
000.0
100.0
Gas Flow(m/s)
02.00
00.00
20.00
Air Purge(kPa)
No
000.0
100.0
005.0
Alarm
4mA
20mA
Under
2007-4-1
12:00:00
CH1
Fixed
Fixed
CH2
Select “Fixed” by the
key or the
key, and press the
ENT
key.
(
)
ENT
Analog Input
GasPress.(kPa)
Fixed
Fixed
4mA
20mA
00.00
00.00
10.00
Gas Temp.(
℃
)
0027
0000
0500
O2
(vol%)
20.00
00.00
21.00
H2O
(vol%)
0
50.0
000.0
100.0
Gas Flow(m/s)
02.00
00.00
20.00
Air Purge(kPa)
No
000.0
100.0
005.0
Alarm
4mA
20mA
Under
2007-4-1
12:00:00
CH1
Fixed
Fixed
Fixed
Enter the fixed value.
Change the numeric value by the
key or
the
key, and move the digits by the
key. Press the
ENT
key to validate the set input
value.
(
)
ENT