- 54 -
INZ-TN1ZSS-E
6.6 Display/AO setting
Make a setting of “Inst./Avg.” or “Wet/Dry” which displays the measurement value or its external output
value displayed on the “Measurement” screen. In case of 2 range specification, range setting is performed.
Explanation of setting items
“Inst./Avg.”
: Make a setting of “instantaneous value” or “average value” for each measur-
able component.
“Wet/Dry”
: Make a setting of “Wet” or “Dry” for each measurable component. When
“Dry” is set, H
2
O setting of the “Analog Input” is required.
“Range Control” : In case of 2 range specification, the following setting can be selected.
“Auto”
: Automatic range switching
“Range1”,
“Range2” : Fixed range
“Remote”
:
Analog
input
switching
Select “Display/AOSetting” from the “Menu” screen, and press the
ENT
key.
2006-04-01
12:00:00
Menu
Zero Calibration
Alarm Log
Display
/
Output
Password
*****
Parameter Setting
Span Calibration
Alarm Setting
Output Hold
AO Select
Input Signal
ENT
6.6.1 Setting of instantaneous/average value
Make a setting of “Instantaneous value” or “Average value” for each measurable component. The setting is
reflected on the display on the “Measurement” screen or external output such as analog output.
O
2
conversion value can be arbitrarily selected even if the previous measurement value is set as instantane-
ous value (or average value).
When the average value is set, moving average processing is performed at the time set at “Average Period”
on the “Parameter Setting” screen.
Display/AOSetting
Range1 0
~
50.0 ppm
Range2 0
~
1000ppm
Range Control
Remote
Inst.
Inst.
Dry
HC
ℓ
cv
2007-
4
-1
12:00:00
Gas
HC
ℓ
Avg.
Inst.
Wet
Wet
Dry
Range1 0
~
50.0 ppm
Range2 0
~
1000ppm