53
D/A Converter
What is "D/A Converter"?
This function allows you to set analog output that links to the display value or analog
output of a load applied to the load cell.
It can change over data to output. See Page 47.
D/A output circuit and the circuit of IN-100 proper are insulated.
You can select the analog output range between 0 ± 10 V by V.
The current is from 4 to 20 mA. No minus output is allowed.
The conversion speed is 4,000 Hz.
D/A Zero
outputs 0 V or 4 mA when the load is the set value.
D/A Full Scale
outputs voltage as set by
D/A Max. Voltage
or 20 mA when the load is
the set value.
Select either
Voltage
or
Current
for analog output. They cannot be output
simultaneously.
D/A Max. Voltage
is the maximum value of the analog output voltage. However, when a
load exceeded the load set by
D/A Full Scale
, the function outputs 10% larger than
D/A
Max. Voltage
and then outputs
DA OVER
message.
How to set
1. Press MENU key.
2. Use
⋀
⋁
key to select
System Setting
.
3. Press SET key.
4. Use
⋀
⋁
key to select
D/A Converter
.
5. Press SET key.
6. Set
D/A Zero
setting. A value when the measured value is zero will be changed.
7. Use
<
>
key to change the digit and
⋀
⋁
key to change the value.
(default: +00000)
8. Press SET key and confirm the value has turned green.
9. Press SET key again to set
D/A Full Scale
.