Sample Extractor Assemblies
FTS Systems LyoStar™ 3
34
Rev 006, 01/14
© SP Scientific 2014
Sample Extractor Operation
To operate the Sample Extractor Assembly:
Ensure that the auxiliary vacuum pump is connected to vacuum pump inlet.
1.
Activate the auxiliary vacuum pump and open the antechamber vacuum valve
2.
(by turning it to the right) to pull vacuum in the antechamber.
When the vacuum gauge reads approximately -30 inHg, close the
3.
antechamber vacuum valve (by turning it to the left) and deactivate the
auxiliary vacuum pump.
Turn the antechamber lock / handle so it is parallel to the sample extractor
4.
(unlocked).
Note: Unlocking the antechamber may cause the lyophilizer’s primary vacuum reading
to increase. The system’s vacuum pump may activate to compensate for the sudden
change in pressure.
Vacuum Pump Inlet
Antechamber
Vacuum Valve
Antechamber Lock / Handle (Unlocked)
Vacuum Gauge
-30 in.Hg
Summary of Contents for LYOSTAR 3
Page 2: ......
Page 6: ...iv Rev 006 01 14 SP Scientific 2014...
Page 10: ...Contents FTS Systems LyoStar 3 viii Rev 006 01 14 SP Scientific 2014...
Page 18: ...Getting Started FTS Systems LyoStar 3 8 Rev 006 01 14 SP Scientific 2014...
Page 30: ...Operating your Lyophilizer FTS Systems LyoStar 3 20 Rev 006 01 14 SP Scientific 2014...
Page 36: ...Liquid Nitrogen Trap FTS Systems LyoStar 3 26 Rev 006 01 14 SP Scientific 2014...
Page 48: ...Sample Extractor Assemblies FTS Systems LyoStar 3 38 Rev 006 01 14 SP Scientific 2014...
Page 52: ...Stoppering FTS Systems LyoStar 3 42 Rev 006 01 14 SP Scientific 2014...
Page 56: ...Shelf Latching FTS Systems LyoStar 3 46 Rev 006 01 14 SP Scientific 2014...
Page 74: ...Appendix A Sample Recipe FTS Systems LyoStar 3 64 Rev 006 01 14 SP Scientific 2014...
Page 80: ...Appendix C Stainless Steel Cleaners FTS Systems LyoStar 3 70 Rev 006 01 14 SP Scientific 2014...
Page 81: ......