FTS Systems LyoStar™ 3
Sample Extractor Assemblies
Rev 006, 01/14
29
© SP Scientific 2014
Align the Sample Extractor Assembly door with the hinge brackets and
3.
reinstall the quick release pins.
Reinstall the front antechamber plate with the selector arm using the
4.
thumbscrews.
Sample Extractor Operation
Before beginning operation of the Standard Sample Extractor Assembly, ensure that
the front plate is closed and that the inert gas/vacuum break valve is closed.
To operate the Sample Extractor Assembly:
Activate the auxiliary vacuum pump to pull vacuum in the antechamber.
1.
After the vacuum gauge stabilizes at the maximum position, allow the vacuum
2.
pump to operate for an additional 15 minutes.
Open the antechamber isolation door. With locking handle in the up position
3.
(unlocked), slide the antechamber door open.
When the antechamber isolation door is fully open, lower the handle to lock
4.
the isolation door in position.
Raise handle to
unlock the door.
Antechamber
Isolation Door
Sample Extractor Assembly Door Without
Front Antechamber Cover And Selector Arm.
Summary of Contents for LYOSTAR 3
Page 2: ......
Page 6: ...iv Rev 006 01 14 SP Scientific 2014...
Page 10: ...Contents FTS Systems LyoStar 3 viii Rev 006 01 14 SP Scientific 2014...
Page 18: ...Getting Started FTS Systems LyoStar 3 8 Rev 006 01 14 SP Scientific 2014...
Page 30: ...Operating your Lyophilizer FTS Systems LyoStar 3 20 Rev 006 01 14 SP Scientific 2014...
Page 36: ...Liquid Nitrogen Trap FTS Systems LyoStar 3 26 Rev 006 01 14 SP Scientific 2014...
Page 48: ...Sample Extractor Assemblies FTS Systems LyoStar 3 38 Rev 006 01 14 SP Scientific 2014...
Page 52: ...Stoppering FTS Systems LyoStar 3 42 Rev 006 01 14 SP Scientific 2014...
Page 56: ...Shelf Latching FTS Systems LyoStar 3 46 Rev 006 01 14 SP Scientific 2014...
Page 74: ...Appendix A Sample Recipe FTS Systems LyoStar 3 64 Rev 006 01 14 SP Scientific 2014...
Page 80: ...Appendix C Stainless Steel Cleaners FTS Systems LyoStar 3 70 Rev 006 01 14 SP Scientific 2014...
Page 81: ......